Upgrade HS-2000 plus

HS-2000 plus

The Sensitivity Enhancement for SCIEX API 2000 TM


The state-of-the-art sensitivity enhancement for your SCIEX API 2000 TM .

HS-2000 plus introduces a 100% efficient collision cell that brings the SCIEX API 2000 performance to the SCIEX API 3000 and 4000 standards and increases the sensitivity in MS/MS of up to four times.

• On-site upgrade with minimum downtime
• One year warranty on upgrade components
• No new training of staff is required
• Guaranteed GLP compliance
• Long-term service contracts

Data are acquired using a Sciex API 2000 TM and HS-2000 plus modified instrument.

Standard solutions are sampled by infusion using nebulization-assisted electrospray ionization.

The sensitivity increase for positive mode (Reserpine) and negative mode (Taurochilic Acid) is illustrated here before (red) and after the upgrade (blue) for Reserpine (left) and T-Acid (right):

 

COMPARING THE Q1 AND Q3 SCANS of the SCIEX API 2000 TM with HS-2000 plus


HS-2000 plus system modification of the SCIEX API 2000 brings its collision cell up to SCIEX API 3000 and 4000 performance standards, making it 100% efficient.

According to SCIEX API 2000 specs, the Q3 scan for positive PPG tests for mass m/z 906, the Q3/Q1 efficiency is 40%. With HS-2000 plus, the efficiency is 100% (Q1 single scan gives 3.0e6 cps and Q3 single scan gives 3.1e6 cps).

For negative PPG tests, the Q3/Q1 efficiency for mass m/z 934 is 20%. With HS-2000 plus, the efficiency is more than 100% (for example, Q1 ten scans gives 1.1e6 cps and Q3 ten scans result in 1.3e6 cps). With high masses the improvement is even better. For the mass m/z 1746 the Q3/Q1 efficiency of a SCIEX API 2000 is 10%, and now with 2.8e5 cps in Q1 scan, HS-2000 plus gives 3.5e5 cps in Q3 scan.

The graphs here illustrate the positive PPG Q1 and Q3 scans efficiency for mass m/z 906 of the SCIEX API 2000 (left) and the HS-2000 plus system (right).